Where are the sense lines connected
Connection of sensors with measuring amplifiers
Sensors with sense wires
If the measuring amplifier has only 4 wires for the sensor, you can
- leave the the sense wires unconnected,
 - connect the sense wires in parallel with the excitation wires.
 
The parallel connection of sense wires and excitation wires is better, because
- The line resistance for excitation wires is reduced,
 - The interference immunity is increased (no open lines).
 
Sensors without sense wires
If the measuring amplifier has a sense wire input, you must install a connection between
- positive excitation and positive sense wire and
 - negative excitation and negative sense wire.
 
For all measuring amplifiers "GSV" of ME-Meßsysteme GmbH, the connection may be omitted, since internallya high-resistance connection is already given.